TION 
                 
              Finger photoplethismography 
                (PPG) is a commonly used technique in medical services. In most 
                of the cases it is used for measuring the heart rate. A recent 
                research suggests the possibility to use plethismography for measuring 
                blood pressure and other indices of peripheral vascular activity 
                [2,4,14,15,17]. 
                However, it has been pointed out that there may be difficulties 
                related to wrong data interpretation [3,6]. 
                
              Nevertheless, 
                several groups suggest that the plethismographic signal carries 
                very rich information about cardiovascular regulation, which commonly 
                is not obtained by using the available methods [1, 
                13,16]. 
              Thus it seems 
                to be justified the use of advanced signal analysis techniques 
                for the study of plethismographic signals. 
              In this work, 
                we applied a high-degree polynomial detrendening technique in 
                combination with a kernel nonparametric method for evaluating 
                plethismographic signal dynamics. 
              Our results 
                indicate that the detrendened plethismographic signal is a very 
                stable, highly nonlinear signal with a very small stochastic contribution. 
                Application of the kernel nonparametric estimator revealed that 
                the purely deterministic nonlinear component corresponds to a 
                low dimensional limit cycle resembling the original appearance 
                of the phase portrait from original data. The stochastic component 
                of this signal, however, is not a white noise. At least an important 
                contribution of 1/f noise may be detected. Regarding its spectral 
                properties, the noise component shares some of the properties 
                of heart rate variability signals reported in literature.
                Thus, the application of nonlinear techniques allowed separating 
                three distinct components of the plethismographic signal: 
               A slow non-stationary 
                component probably related to recording artifacts. 
               A non-linear 
                deterministic component carrying information about the waveform 
                features, and corresponding to a limit cycle. 
              A noise non-stationary 
                component with 1/f dynamics, suggesting at least partial fractality. 
                This component seems to share important aspects of the R-R signal 
                described in literature. 
              The two last 
                components may carry important information both, about the cardiovascular 
                and the autonomic nervous system. Thus, we may expect that the 
                previously mentioned methodology may open new possibilities both 
                for research as well as for diagnostic purposes. 
                
              
              
               MATERIALS 
                AND METHODS 
              Subjects 
                
              Transmission 
                plethismography (PG) recordings were obtained from six healthy 
                volunteers (three female and three males), aged 24-46 years (average 
                age 32 years). The subjects were seated during examination, with 
                their hands laid comfortably on the table at the level of the 
                heart. Recording time was 5 min after a rest period of 10-min. 
                Room temperature was 25 oC. 
              PPG Recordings 
                
              The transmission 
                PPG probe of the commercial pulse oxymeter (OXY9800-COMBIOMEQ, 
                Cuba) consisted of a light emitting diode (LED) of 865 nm and 
                a PIN photodetector (peak spectral response 865 nm) placed on 
                different sides of the probe, so that they were attached to the 
                two contralateral surfaces of the finger. 
              The PPG probe 
                was attached to the right hand. The computerized system allowed 
                to optimally adapting the amplification level for obtaining a 
                nonsaturated signal. The PPG signal was digitized (75 Hz) and 
                stored in ASCII format. All the statistical processing was performed 
                off-line. 
              Data 
                analysis. 
              Signal processing 
                included: 
              
                -  Power 
                  spectrum estimation. 
- Baseline 
                  correction. 
- Correlation, 
                  dimension, estimation. 
- Fractal 
                  Dimension estimation. 
- Nonlinear 
                  Dynamics identification. 
Power 
                spectrum estimation was performed through the FFT 
                algorithm applied to the original signal [20]. The power spectrum 
                was estimated as the square of the absolute value of the FFT. 
                Linear fit was performed to log-log transformed power spectra 
                for obtaining an estimate of the fractality index a 
               Baseline 
                correction was performed by adjusting the whole 
                signal to a polynomial of the type:
                
                where St is the plethismografic signal evaluated at 
                time t; a0,a1,…,ak, are 
                real constants, corresponding to the model’s coefficients. 
                The order of the model was set at k=22, since this value better 
                warranties stationarity of the corrected signal without affecting 
                the shape of the pulsatile waveforms. 
              The aim of 
                this procedure is to correct the signal for nonstationarities, 
                while preserving the original features of the waveform. 
              Three experts 
                independently checked for the preservation of the waveform’s 
                shape after the detrendening procedure. 
              Correlation 
                dimension estimation a Grassberger Procaccia algorithm 
                [10], was applied to the detrendened signals. 
                In each case the estimated value was compared with that of a surrogate 
                signal obtained via inverse FFT analysis with phase randomization, 
                as described in [10]. Two-dimensional projections 
                of the reconstructed phase portraits were obtained applying the 
                Taken’s vector method [8]. 
              Nonlinear 
                Dynamics identification. Kernel nonparametric analysis 
                was applied to the trend-corrected signals. In kernel autoregression, 
                the signal is fitted to a model of the type: 
                 
 
                
                The function nonlinear F is obtained as a weighted average of 
                the observed points in the phase space, the nearest points bearing 
                the highest contribution. A detailed description of the method 
                appears in the references [7-11, 
                18]. 
              During the 
                application of the kernel procedure, the following information 
                was obtained: 
              
                - Order of 
                  the autoregressive model (r). It reflects the number of past 
                  values necessary to optimally describe the autoregressive function. 
                  
   
  Nonlinear 
                correlation coefficient expressed as: 
              
                -  where 
                  Vtot corresponds to the signal’s variance and Vne is the 
                  unexplained variance after applying the model. 
- In the 
                  linear case, this expression is equivalent to the linear correlation 
                  coefficient [10]. 
- Noise free 
                  realization generation (NFR). The noise free realization [9,18] 
                  is obtained via sequential estimation of the function F to previously 
                  estimated data points of the time series. The first points of 
                  the series are set at random. The 100 first points of the NFR 
                  are discarded for assuring the absence of transients in the 
                  NFR. 
The phase 
                portrait reconstructed from the NFR gives information about the 
                noise free dynamical system. 
                
              
              
              
                RESULTS 
              General features 
                of the signal. 
              An original, 
                untreated recording is shown in fig 1. 
              
                Fig 1. Photoplethismographic (PPG) signal obtained from a female 
                healthy subject (age 30 years). Legend: Abscissas-time, expressed 
                in 13,33-ms time units. Ordinates- PPG signal amplitude in arbitrary 
                units. 
              As appreciable, 
                significant baseline shifts are present in this signal. It is 
                possible that this is a consequence of recording difficulties 
                (fixation fails, subject’s movements, etc), though it is 
                not possible to exclude some physiological influences. A detailed 
                look up at the signal shows the presence of quite stable waveforms, 
                whose appearance is relatively little influenced by baseline shifts. 
                
              In the literature, 
                most of the attention is paid to pulsatile waveforms ([1,13,19]). 
                
              It could be 
                plausible to suppose that this signal is a more or less periodic 
                component affected by random influences. However, some evidences 
                may not support this view. 
              In fig 2 a 
                log-log plot of the signal’s power spectrum is shown. As 
                appreciable, a linear component with a negative slope could be 
                appreciated. The estimated slope from the power spectrum is 1.83, 
                close to that of fractal Brownian process [12]. 
                
               
 
              
              Fig 2. Log-Log 
                plot of the power spectrum estimated from the recording in 1-a. 
                Legend: Abscissas, frequency in arbitrary units; Ordinates-Power 
                spectral density. Notice the presence of a relevant negative slow 
                linear component. The peak corresponds to the periodic pulsatile 
                waves. 
              We applied 
                a time domain algorithm proposed by Higuchi for the estimation 
                of the fractal dimension [12]. Higuchi's method 
                also allows the estimation of the power-spectral slope in log-log 
                coordinates (a). 
              The signal’s 
                dimension was 1.65, which correspond to a fractal process. The 
                spectral log-log slope (a) estimated by Higuchi’s method 
                was 1.70, close to that estimated from the power spectrum (1.83, 
                see above). 
              Since these 
                observed features of the original signal may not be explained 
                by its periodic component, nor by its perturbations with Gaussian 
                noise, it has sense to try to find out what components may be 
                responsible for different properties of the PG signal. 
              Base Line 
                Correction 
              Fig 2a. shows 
                an original signal and the estimated 22-degree-polynomial. As 
                appreciable, the procedure applied gives a satisfactory estimate 
                for a time-dependent base line for the original signal. 
               
 
              
              Fig 2-a. The 
                recording from fig in 1-a and its time-dependent trend estimated 
                by fitting to a 22-degree polynomial. Legend: as in fig 1a 
              After base 
                line substraction, a detrendened signal was obtained (fig 2b). 
                
               
 
              
              Fig 2-b. The 
                result of subtracting the estimated trend in fig 2a from the signal 
                in fig 2. This is the trend-corrected component. Legend: as in 
                fig 2. 
              Visual appreciation 
                suggests that the detrendened signal seems to be much more close 
                to a stationary one than the original signal. This may also be 
                supported with quantitative data. Fig 2c represents the dependence 
                of the standard deviation of the signal for both, the original 
                and the detrendened signal. This dependence for the original signal 
                is typical for nonstationary time series, included fractal ones 
                ([12,20]). 
              
                Fig 2-c. Dependence of the signal’s standard deviation upon 
                segment’s length for the original trace in fig 2 (upper 
                line) and for the trend-corrected signal from fig 2b. Legend: 
                abscissas-time-segment duration in 13.3 ms sampling units. Ordinates: 
                standard deviation. 
              At the same 
                time, the detrendening procedure did not affect the appearance 
                of the pulse waveforms. 
              Fig. 3a shows 
                a 2-dimensional projection of the reconstructed phase portrait. 
                
               
 
              
              Fig 3a. S-D 
                projection of the reconstructed attractor, using the Takens method. 
                Abscissa St. Ordinates St-10.This picture corresponds to the trend-corrected 
                signal in fig 2b. 
              In figure 
                3b is represented the same phase portrait obtained from another 
                record taken from the same subject 3 days after. The high stability 
                of the picture may be suggested from both pictures. 
               
 
              
              Fig 3b. The 
                same as in fig 3a, applied to another recording from the same 
                subject taken three days after. Notice that the difference in 
                scales accounts for very similar attractor geometry. 
              The appearance 
                of this signal could suggest about a low dimensional chaotic dynamics, 
                very similar to the Rossler attractor. 
              However, as 
                shown previously for EEG signals, an alternative explanation may 
                be that of a limit cycle perturbed by noise [9]. 
              We estimated 
                the correlation dimension of the detrendened signal. For that 
                we applied the Grassberger-Procaccia algorithm. The value of correlation 
                dimension corresponding to the detrendened signal was D=4.694 
                ± 0.662 for an embedding dimension of ED=10. This value 
                may suggest about a low dimensional chaotic attractor. However, 
                for the phase surrogate of the detrendened signal this value was 
                3.64 ± 0.51. Thus correlation dimension value does not 
                support the hypothesis of a chaotic attractor. 
              Nonlinear 
                identification techniques allow to separate deterministic and 
                stochastic components from a nonlinear stochastic time series. 
                One of the most powerful methods of nonlinear identification is 
                the kernel nonparametric autoregressive estimation. However, this 
                method works satisfactorily only for stationary signals. Thus 
                the detrendening procedure allowed us to apply a nonlinear identification 
                method to detrendened data. 
              Application 
                of this method to 1000-points segments of the original time series 
                revealed that the order of the autoregressive model was equal 
                to 3 in more than 85% of the segments analyzed. The nonlinear 
                correlation coefficient was higher than 0.99 in all cases, suggesting 
                the presence of a low-dimensional nonlinear deterministic component 
                in the detrendened signal. 
              Fig. 4 shows 
                a phase portrait obtained from the noise-free realization. The 
                comparison to figs. 3a and 3b supports the idea of the plethismographic 
                signal modeled as a limit cycle perturbed by random contributions. 
                
               
 
              
              Fig 4. The 
                same as in fig 3a, applied to a noise-free realization obtained 
                from applying kernel nonparametric autoregression to the first 
                1000 points of the trend-corrected recording in fig 2b. Notice 
                the limit cycle structure of the noise-free attractor. 
              Though the 
                generated noise free realization may carry information about the 
                nonlinear dynamics of pulse wave generation, it obviously may 
                not account for the fractal-like properties of the original recording. 
                
              Properties 
                of the Residuals 
              From the application 
                of the kernel autoregression there still remain the residuals 
                of the estimation. The residual’s signal may give a good 
                estimate of the noise component in the model. If we compare the 
                variance of the noise component to that of the original signal 
                we may observe that this component seems to be negligible compared 
                to the original signal or even to the noise free realization (it’s 
                value never reaches 5% of the variance of the detrendened signal). 
                
              However, the 
                residual’s signal also may contain interesting information. 
                
              In particular, 
                the residuals’ signal (fig 5a) does not seem to be a white 
                noise signal. Is spectral slope in log-log coordinates is a=1.3 
                (r=0.73; n=450). Though it is lower than the value of the original 
                signal this suggests about the presence of fractal components 
                in the residual’s signal (see fig 5b). 
               
 
              
              Fig 5a Residuals 
                after the estimation of the expected values in the recording from 
                fig 2b. according to the nonlinear nonparametric autoregressive 
                model. Legend, as for fig 2. Notice the small amplitude of the 
                signal, compared to fig 2b, or 2a. 
               
 
              
              Fig 5b. Log-log 
                power spectrum of the residual’s signal in fig 5. Legend, 
                as in fig 2. After discarding the first 15 points on the left 
                part of the trace, the curve fits to a straight line with a = 
                0.83 and the linear correlation coefficient r=0.76 (N=450 data 
                points). 
              Finally, we 
                decided to reconstruct a signal composed by the sum of both residuals 
                and the estimated base line. This signal will carry information 
                about the original signal with exception of the part corresponding 
                to a deterministic nonlinear dynamics. 
              After submitting 
                this signal to the f Higuchi’s algorithm, the estimated 
                signal’s fractal dimension was D=1.65, which corresponds 
                to a value of a=1.71, almost identical to the 1.70 value obtained 
                from the original data (see above). 
              Fig. 6 represents 
                the log-log spectral plot of the baseline-added residual signal. 
                
               
 
              
                
              Fig 6- The 
                log-log power spectrum of the sum of the signal in 5 and the trend 
                component from fig 2a. Similarity to fig 2 is appreciable. 
              
              
              DISCUSSION 
                
              The main results 
                obtained in this research may be summarized as follows: 
              The PPG signal 
                can be represented as a sum of at least three processes, including: 
                
              
                - A low dimensional 
                  nonlinear component with a periodic attractor, which accounts 
                  for the plethismographic signal waveform, commonly studied [13]. 
                  
-  A more 
                  or less regular high-amplitude-baseline signal, which may be, 
                  approximates as a polynomial function of the time. The baseline 
                  is likely to include both artifacts (patient’s movements, 
                  sensor’s fluctuations, etc) and physiological components 
                  (tissue replenishing with blood, respiration, etc. [16]). 
                  
- A low-amplitude 
                  random component with fractal-like properties. 
The original 
                signals present not only periodicities due to the presence of 
                the pulse waves, but also fractal-like properties. The sum of 
                the baseline and the residual’s signal (after applying a 
                kernel nonparametric method) may account for the fractal-like 
                properties of the original signal. 
              Thus our results 
                suggest that it is possible to separate the plethismographic signal, 
                via nonlinear filtering, into components carrying different types 
                of information. 
              In our opinion, 
                our results open new possibilities, and, at the same time, give 
                birth to new questions. 
              Thus the fact 
                that the nonlinear deterministic part of the signal may be modeled 
                with a nonparametric function of order 3 opens the possibility 
                to find an analytical model for this component generation. Since 
                this component is related to arterial compliance and resistance, 
                as well as other factors contributing to blood volume increase 
                during systole [16], this may open new possibilities 
                for research and diagnosis of blood pressure regulation. 
              It seems difficult 
                to determine the possible physiological components of the baseline 
                signal. Though literature data refer to the baseline as inversely 
                related to the blood volume in the tissue under examination [16], 
                it is not excluded the contribution of recording artifacts to 
                this signal. Thus, as described in [1] minor 
                movements of the arm may change the signal baseline. It seems 
                reasonable to suppose that physiologically-conditioned factors 
                contribute to the fractality of the original signal, which could 
                not be expected from low frequent trends in the signal. 
              At the same 
                time, it seems difficult to ignore the physiological implications 
                of the fractality of the low amplitude residual (noise) signal. 
                
              A considerable 
                experimental material supports the idea of a partially fractal 
                nature of heart rate variability signals [5,20]. 
                
              However, the 
                HRV signal is contained in the electrocardiogram, and, as assumed 
                also in the plethismographic signal [16]. 
              The signals 
                analyzed were relatively short in terms of HRV analysis, and lasted 
                fractions of a minute. However, if we believe in the truly fractality 
                of a process, it must be present not only at long-range time scale, 
                but also at microscopic scales [5]. 
              On the other 
                hand different authors have claimed the fractal nature of ECG 
                complexes [5]. Thus it would not be surprising 
                to find a physiologically conditioned fractal component in the 
                PPG signal. 
              A major task 
                remains to identify the physiological bases of these processes. 
                In literature, several attempts have been made, as for example 
                those related to interpret the fractal nature of the QRS-complex 
                in the ECG as emerging from the fractal nature of the Hiss bundle 
                [5]. 
              Experimental 
                research by Yamamoto suggested that the HRV fractal component 
                is not affected by beta blocking, which suggests involvement of 
                factors different from the autonomic nervous system in this process 
                [20]. 
              The possible 
                diagnostic utility of these results also open new perspectives. 
                Some of these questions are subject of further analysis by our 
                group. 
              
              
               REFERENCES 
                
              1. 
                Allen J, Murray modeling the relationship between peripheral blood 
                pressure and blood volume pulses using linear and neural network 
                system identification techniques. A Physiol Meas 1999 Aug;20(3):287-30.
                2. Bruner, J. M. R. (1981). Comparison of direct 
                and indirect methods of measuring arterial blood pressure. Medical 
                Instrumentation, 15, 11-12.
                3. Escourrou PJ, Delaperche MF, and Visseaux A, 
                "Reliability of Pulse Oxymetry during Exercise in Pulmonary 
                Patients", Chest, 1990, 97 (3): 635-8.
                4. Frey, W. Butt, Neonatal and pediatric intensive 
                care: Pulse oxymetry for assessment of pulsus paradoxus: a clinical 
                study in children, Intensive Care Medicine 24 (1998), 242-246.
                5. Goldberger, A.L. Fractal Mechanisms in the 
                electrophysiology of the heart. IEEE Transactions on Engineering 
                in Medicine and Biology (11): 47-52, 1992.
                6. Gregorini P, Gallina A, and Caporaloni M: Comparison 
                of One Minute versus Five Minute Sampling Rate of Physiologic 
                Data. The Internet Journal of Anesthesiology 1997; Vol1N4: 
                http://www.ispub.com/journals/IJA/Vol1N4/sampling.htm.
                7. Haerdle W, Luetkepohl H, Chen R. A review of 
                nonparametric time series analysis Int Stat Rev. 1997 65 1: 49-72.
                8. J. L. Hernández, L. García, Guido 
                Enzmann, A. García. “La regulación autonómica 
                del intervalo cardíaco modelada como un sistema no lineal 
                estocástico con múltiples atractores”. Revista 
                CENIC. Ciencias Biológicas. Vol. 30, No 3, 1999.
                9. J. L. Hernández, P. Valdés and 
                P. Vila. "Spike and wave activity with a limit cycle perturbed 
                by noise". NeuroReport, Vol. 13, 1996.
                10. Hernández JL, Valdés JL, Biscay 
                R, Jiménez JC, Valdés P. "EEG predictability. 
                Adequacy of nonlinear forecasting methods". Int J Biomed 
                Comput. 38 197-206 (1995).
                11. Hernández, J. L., Biscay, R, Jiménez, 
                JC, Valdés P, Grave de Peralta, R. "Measuring the 
                dissimilarity between EEG recordings through a non linear dynamical 
                system approach". Int J. Biomed. Comput. 38, 121-125 (1995).
                12. Higuchi Physica D, 1990.
                13. Imanaga I, Hara H, Koyanagi S, Tanaka K Correlation 
                between wave components of the second derivative of plethismogram 
                and arterial distensibility. Jpn Heart J 1998 Nov;39(6):775-84.
                14. Jennings, J. R., Tahmoush, A. J., & Redmond, 
                D.P. (1980). Non-invasive measurement of peripheral vascular activity. 
                In I. Martin & P. H. Venables (Eds.), Techniques in Psychophysiology. 
                John Wiley & Sons, Ltd.: New York.
                15. Mineo R, Sharrock NE Pulse oximeter waveforms 
                from the finger and toe during lumbar epidural anesthesia. Reg 
                Anesth 1993 Mar-Apr;18(2):106-9.
                16. M. Nitzan, A. Babchenko, B. Khanokh. Very 
                low variability in arterial blood pressure and blood volume pulse. 
                Med. Biol. Eng. Comput., 1999, 37, 54-58.
                17. Rumwell C.,McPharlin M: In Vascular Technology, 
                an illustrated review for the registry exam. Davies Publishing 
                Inc., Pasadena, CA, 1996.
                18. P. Valdés, J. Bosch, J. C. Jiménez, 
                N. Trujillo, R. Biscay, F. Morales, J. L. Hernández, T. 
                Ozaki. The statistical identification of nonlinear brain dynamics: 
                A progress report. In: “Nonlinear Dynamics and Brain Functioning”. 
                Pradhan N., Rapp P. E. And Sreenivasan (Eds.), Nova Science Publishing, 
                1999.
                19. M. Wolf Pulse Oxymetry and Medical Infrared 
                Spectroscopy http://www.biomed.ee.ethz.ch/staff/ibt_members.html 
                (1998).
                20. Yamamoto, Y., Hughson, R.L. Coarse graining 
                spectral analysis: new method for studying heart rate variability. 
                Journal of Applied Physiology (71): 1143-1150,1991.